Clinical MeasurementGas analysis and capnography
01

The physical principle

Why some gases absorb infrared and others cannot

Every clinical carbon dioxide analyser in an operating theatre works by infrared absorption. Understanding one property of the molecule tells you which gases can be measured this way, at what wavelength, and why oxygen needs a different technique entirely.
Estimated study time

About 80 minutes

Plus the time it takes to redraw this lesson’s figures from memory, which is the fastest way to find out what you have not understood.

Why it matters

Where this shows up

Capnography is the monitor that confirms the tube is in the trachea, and there is no substitute for it in that moment. This lesson covers how the measurement is actually made, every way the reading can be wrong, and what each shape of the trace means.

Learning outcomes

By the end of this lesson you should be able to:

  1. State which gases absorb infrared and which cannot, and explain the molecular property that decides it.
  2. State the Beer–Lambert relationship with every symbol defined, and separate its two component laws.
  3. Describe how a capnometer measures carbon dioxide, and distinguish that from how gas is delivered to the analyser.
  4. Compare sidestream with mainstream sampling on delay, water handling, dead space and which gases can be measured.
  5. Describe zeroing and calibration, and give each source of error with its mechanism, the direction of the error and its remedy.
  6. Draw a normal capnogram with labelled and scaled axes, four numbered phases, the end-tidal point and the α and β angles.
  7. Interpret abnormal traces, and quote the normal end-tidal value and the arterial to end-tidal gradient in both units.
  8. Describe paramagnetic, fuel cell and polarographic oxygen analysis, and the physiological effects of hypercarbia.

Together these settle 5 syllabus objectives: Infrared absorption and the Beer–Lambert law; Capnometer design, sampling, calibration and error; The capnogram, and the arterial to end-tidal gradient; Paramagnetic, fuel cell and polarographic oxygen analysis and Hypoxia, hyperoxia, hypercapnia, respiratory inadequacy and failure. Tick them on the Clinical measurement objective list once you can do all of the above without notes.

A molecule absorbs infrared radiation only if its vibration changes its dipole moment — that is, only if the distribution of charge across the molecule is asymmetrical or becomes asymmetrical as the bonds stretch and bend. Carbon dioxide, nitrous oxide, water vapour and the halogenated volatile agents all satisfy this. Oxygen, nitrogen and helium do not: they are elementary or symmetrical, have no dipole moment, and are effectively transparent to infrared.

Where each gas absorbs

GasAbsorbs infrared atBand used clinicallyWhy it matters
Carbon dioxide4.26 µmNarrow band, 4.2–4.3 µmThe working wavelength of every theatre capnometer
Nitrous oxide4.5 µm (also absorbs near 4.3 µm)Separate bandSpectral overlap and collision broadening both cause CO₂ over-reading
Volatile agents3.3 µm (C–H bond stretching)3.3 µm, with several bands used for agent identificationAll the modern agents share this band, so identification needs more than one wavelength
Water vapourBroad absorption across the infraredRemoved rather than measuredCondensation and vapour are a source of error, not a signal
Oxygen, nitrogen, heliumNoneNot measurable by infraredElementary and symmetrical: no dipole moment, so no infrared absorption

The Beer–Lambert law

Infrared light of the chosen wavelength is passed through a chamber of known path length containing the sample. The transmitted intensity falls exponentially with both the concentration of the absorbing gas and the distance the light travels through it.

Beer–Lambert lawI = I₀ · e−ε·c·l
  • I — transmitted intensity reaching the detector
  • I₀ — incident intensity leaving the source
  • ε — the extinction coefficient (molar absorptivity): how strongly one mole of this particular gas absorbs at this particular wavelength, in L·mol⁻¹·cm⁻¹. It is a fixed property of the molecule, not of the sample or the instrument, and it is the term that makes the method gas-specific — carbon dioxide has a large ε at 4.26 µm and a negligible one elsewhere, which is exactly why a narrow optical filter can pick it out of a mixture. A gas with ε = 0 at the chosen wavelength, such as oxygen, cannot be measured this way at all.
  • c — concentration of the absorbing gas, the quantity being measured
  • l — path length through the sample chamber, fixed by the design

The two component laws are worth separating, because each states something different. Beer's law states that absorption is proportional to the concentration of the absorbing substance. Lambert's law states that absorption is proportional to the path length. Combined, the energy absorbed from the narrow infrared band passing through the chamber is proportional to the number of absorbing carbon dioxide molecules present. Concentration is then obtained by comparing the measured absorbance with that of a known calibration standard.

02

Instrument design

Inside the capnometer, and where you attach it

The optical bench is the same in every design. What differs is where the gas is analysed: drawn away to a remote analyser, or measured in the breathing system at the airway.

Components of the optical bench

  1. Infrared source — a heated element emitting a broad infrared spectrum.
  2. Optical filter, or a rotating filter wheel, selecting the narrow 4.2–4.3 µm band. Selectivity here determines how much nitrous oxide interference gets through.
  3. Sample chamber of fixed path length, with sapphire windows. Ordinary glass cannot be used because glass itself absorbs infrared.
  4. Reference path or reference cell, containing a gas that does not absorb at the chosen wavelength, so the analyser measures a ratio rather than an absolute intensity and is insensitive to drift in the source.
  5. Photodetector and processor, converting transmitted intensity to a displayed concentration or partial pressure.
Sourceheated elementbroad infraredChopperalternatesthe two pathsfilter 4.2–4.3 µmSample chamberfixed path length lsapphire windowsgas inoutReference chambernon-absorbing gassame path lengthattenuatedunattenuatedPhotodetectorProcessor38 mmHg(5.1 kPa)ratio → Beer–LambertGlass absorbs infrared, so the windows are sapphire. Water vapour absorbs broadly and is removed before the chamber rather than measured.Zeroing against carbon dioxide-free gas fixes I₀ and corrects the offset; calibration against a known mixture fixes the gain.Nitrous oxide broadens the carbon dioxide peak by molecular collision and absorbs nearby, so an uncorrected analyser over-reads.The reference path is why the output is a ratio: drift in the source affects both beams equally and cancels.
Original teaching diagramThe infrared bench of a capnometer. Name the components in this order in an answer — source, chopper, filter, chamber, detector — and the description writes itself. The reference path is what makes the reading a ratio, so drift in the source cancels out instead of appearing as a change in carbon dioxide.

Sidestream against mainstream

SidestreamMainstream
Where the analyser sitsRemote from the patient; gas is aspirated to itIn the breathing system at the airway
Sampling50–200 mL/min drawn through fine-bore tubingNo gas removed; the chamber is part of the circuit
DelayTransit delay of roughly 2–4 s, plus the analyser rise timeEffectively real time
Weight at the airwayLight — only the sampling lineHeavier cuvette, with drag and disconnection risk
Water and secretionsWater trap and filter required; the line can block or kinkCuvette heated to about 40 °C to prevent condensation; windows can be soiled
Dead spaceNegligible added dead spaceAdds apparatus dead space, which matters in neonates
Other gasesThe same sample can be analysed for O₂, N₂O and volatile agentsUsually carbon dioxide alone
Non-intubated patientUsable through nasal cannulae or an oxygen maskNot usable without an airway device
Main drawbackDelay, blockage, gas removal from the circuit in low-flow or paediatric useBulk at the airway, cost of the cuvette, cleaning
Sidestreamthe gas is drawn away to an analyser inside the monitorfine-bore line · 50–200 mL/minPatientcatheter mountbreathing systemwater trapand filterAnalyserinside the monitorCO₂, O₂, N₂O, agentscavenged, orreturned to the circuitTransit delay of 2–4 s plus the analyser rise time · the line can kink, block or leakRemoves gas from the circuit, which matters at low flows and in small children · one sample gives every gasMainstreamthe chamber is part of the breathing system, at the airwayPatientIR sourcedetectorheated cuvette, about 40 °Cbreathing systemelectrical cable only — no gas leaves the circuitMonitordisplay onlyEffectively real time — no transit delayAdds weight and apparatus dead space at theairway, which matters most in neonatesUsually measures carbon dioxide alone · needs an airway device · the windows can be soiled by secretions
Original teaching diagramSidestream against mainstream. The whole distinction is where the gas is analysed — drawn away down a line, or in the circuit at the airway. It is not about portability; monitors of both kinds are portable.
03

Performance and error

Calibration, response time and the ways the reading goes wrong

No account of a measurement is complete without its sources of error. Each one below has a physical mechanism, a predictable direction of error and a specific remedy.

Zeroing and calibration

The analyser is zeroed against a carbon dioxide-free gas, usually room air, which sets the value of I₀ and compensates for gradual soiling of the windows and ageing of the source. It is calibrated against a gas mixture of known carbon dioxide concentration, which sets the scale. Zeroing corrects the offset; calibration corrects the gain. Both are needed.

Response time

The total delay between gas leaving the alveolus and the number changing on the screen is the transit time along the sampling line plus the rise time of the analyser itself, conventionally the time to move from 10% to 90% of a step change. If that total approaches the duration of expiration, the plateau is never reached and end-tidal carbon dioxide is under-read. This is why neonatal and high-respiratory-rate monitoring is the setting where sidestream sampling most often fails.

Sources of error

SourceMechanismEffect on the readingRemedy
Collision (pressure) broadeningNitrous oxide and oxygen broaden the CO₂ absorption peak, so more infrared is absorbed than the CO₂ concentration alone would explainOver-reading of carbon dioxideModern analysers correct automatically once the gas mixture is entered or measured
Overlapping absorption spectraNitrous oxide also absorbs close to the CO₂ bandOver-readingNarrow, well-selected optical filters; multi-wavelength correction
Water vapour and condensationWater absorbs infrared broadly and droplets scatter itUnpredictable error; blocked sampling lineWater trap, hydrophobic filter, heated cuvette, Nafion tubing
Soiled or scratched windowsReduced transmitted intensity is read as absorptionOver-reading and driftCleaning and regular zeroing
Leak or kink in the sampling lineEntrained room air dilutes the sampleUnder-reading and a damped, blunted waveformInspect the line and the connections; check the water trap
Excessive sampling rateFresh gas is entrained during expiration, particularly at small tidal volumesUnder-reading, loss of the plateauReduce the sample flow in paediatric and low-flow use
Slow response or long transit timeRise time plus transit time exceeds the duration of expirationPlateau never reached; under-reading at high respiratory ratesShort, wide-bore sampling line; mainstream sampling in neonates
Barometric pressureThe analyser measures partial pressure but may display volumes per centApparent discrepancy at altitude or under hyperbaric conditionsState which quantity is displayed; pressure compensation

Interactive figure

The capnogram

Select a phase to see what gas is being sampled and why the trace has that shape, then switch to the abnormal traces and read each one as a deviation from the normal waveform held behind it.

Normal trace · ETCO₂ 38 mmHg (5.0 kPa) · 15 breaths per minute · baseline zero
015304560PCO₂ (mmHg)Time (seconds)ETCO₂ 38 mmHgEXPIRATIONINSPIRATION

Four phases, two angles, one end-tidal point

Select a phase above. The marks in this question are labelling marks, so practise naming each part in order: phase I dead-space gas, phase II mixing upstroke, phase III alveolar plateau, the end-tidal point at the end of that plateau, and phase IV the inspiratory downstroke.

04

Reading the waveform

The normal capnogram, phase by phase

The trace carries its meaning in its labelling: both axes with units, a baseline on zero, four numbered phases and the end-tidal point at the end of the plateau. Without those it is a shape, not a measurement.
PhaseWhat it isGas being sampledShape
Phase IStart of expirationCarbon dioxide-free gas from the conducting airways — the anatomical dead spaceFlat, on a zero baseline
Phase IIMixing upstrokeDead-space gas mixing with arriving alveolar gasRapid S-shaped rise
Phase IIIAlveolar plateauCarbon dioxide-rich gas from the alveoliNear-horizontal with a slight upslope; end-tidal carbon dioxide is the value at its end
Phase IVInspiratory downstrokeFresh gas washing the sampling pointTurns through almost a right angle and falls rapidly to baseline

The slight upslope of phase III is real and is worth explaining rather than drawing flat: alveoli with longer time constants and lower ventilation–perfusion ratios empty later in expiration and carry more carbon dioxide, so the concentration creeps upward across the plateau.

The two angles

  • α angle — between phases II and III, normally about 100–110°. It increases with obstruction to expiratory flow and with ventilation–perfusion mismatch.
  • β angle — between phase III and the inspiratory downstroke, normally about 90°. It increases with rebreathing.

Normal values

QuantitykPammHgComment
PaCO₂4.7–6.0 kPa35–45 mmHgArterial partial pressure
PETCO₂4.0–5.3 kPa30–40 mmHgEnd of the alveolar plateau
Pa–ETCO₂ gradient0.3–0.7 kPa2–5 mmHgWidens with alveolar dead space
Mixed venous PCO₂6.1 kPa46 mmHgReference point on the CO₂ dissociation curve
CO₂ production≈ 200 mL/minResting adult; rises with metabolic rate

Why end-tidal carbon dioxide is lower than arterial

The arterial to end-tidal difference exists because alveolar dead space — alveoli that are ventilated but not perfused — contributes carbon dioxide-free gas to the expirate and dilutes the alveolar sample. In a healthy anaesthetised adult the gradient is small and reasonably stable, which is what makes end-tidal carbon dioxide a useful surrogate for arterial carbon dioxide.

It becomes unreliable exactly when the gradient changes. It widens with hypovolaemia, reduced cardiac output, pulmonary embolism, high airway pressures or excessive positive end-expiratory pressure, chronic lung disease, increasing age and the upright posture. In those situations a falling end-tidal value does not mean a falling arterial value — it may mean the opposite.

05

Pattern to mechanism

Abnormal traces, and what capnography is actually for

Work through the interactive figure above before this section. Every abnormal capnogram is read as a deviation from the normal waveform, and each deviation has a mechanism you should be able to state in one sentence.

A diagnostic sequence that works under pressure

  1. Is there a waveform at all? No waveform is an airway or circulatory emergency: disconnection, oesophageal intubation, total obstruction, apnoea or cardiac arrest.
  2. Does the baseline reach zero? A raised baseline is rebreathing — exhausted absorbent, an incompetent valve or inadequate fresh gas flow.
  3. Is there a true plateau? An upsloping, shark-fin expiratory limb with a widened α angle is obstruction to expiratory flow.
  4. Is the height right? Normal shape with a changed height is a ventilation, circulation or production problem — and the direction and speed of change separate them.
  5. Is anything superimposed? A single notch is patient effort; small regular ripples at the cardiac rate are cardiogenic oscillations.

Causes of a raised end-tidal carbon dioxide

Organise these by mechanism rather than listing them at random. The three-way split is what makes the list explanatory, and it stops a single category standing in for the whole.

MechanismCauses
Increased endogenous productionMalignant hyperthermia · thyroid storm · phaeochromocytoma · neuroleptic malignant syndrome · sepsis · fever · shivering · seizures · light anaesthesia · overfeeding, particularly a high carbohydrate load · release of a limb tourniquet or reperfusion of an ischaemic bed · a rise in cardiac output delivering more carbon dioxide to the lungs, including return of spontaneous circulation
Exogenous carbon dioxide loadCarbon dioxide insufflation during laparoscopy, worse with extraperitoneal insufflation · sodium bicarbonate administration · transfusion of stored blood
Reduced eliminationHypoventilation from opioids, volatile agents, residual block, inadequate set minute ventilation, high spinal or neuromuscular disease · increased physiological or apparatus dead space · rebreathing from an exhausted absorber, an incompetent valve or inadequate fresh gas flow · airway obstruction and bronchospasm · reduced functional residual capacity from pneumoperitoneum or head-down positioning · chronic obstructive pulmonary disease
ArtefactCollision broadening by nitrous oxide · an uncalibrated or unzeroed analyser · soiled windows

The clinical roles of capnography

RoleWhat the waveform gives you
Confirming tracheal intubationA sustained waveform over six breaths. The most important single use, and the standard of care.
Continuous airway monitoringDetects disconnection, extubation, obstruction and circuit failure earlier than any other monitor.
Adequacy of ventilationContinuous trend of hypo- and hyperventilation between blood gases.
Adequacy of circulationA fall in end-tidal carbon dioxide with an unchanged ventilation reflects reduced pulmonary blood flow.
Cardiopulmonary resuscitationGuides compression quality and signals return of spontaneous circulation.
Diagnosis of the breathing systemA raised baseline identifies rebreathing; a shark fin identifies obstruction.
Metabolic monitoringA progressive unexplained rise is the earliest sign of malignant hyperthermia.
Depth of block and patient effortA curare cleft signals returning spontaneous respiratory effort.
Sedation and recoveryNasal capnography detects apnoea and airway obstruction earlier than pulse oximetry.
06

Alternatives and their place

Other ways to measure carbon dioxide

These are the alternatives to infrared absorption. None of them is what measures carbon dioxide in a theatre breathing system, and knowing why is the point of listing them.
  • Severinghaus electrode. Carbon dioxide diffuses across a carbon dioxide-permeable membrane into a thin film of bicarbonate solution, where it forms carbonic acid and dissociates. The resulting change in hydrogen ion concentration is measured by a glass pH electrode; a tenfold change in carbon dioxide partial pressure produces approximately one unit of pH change. This is how a blood gas analyser measures PCO₂ in blood, not in gas.
  • Colorimetric detector. A pH-sensitive dye — classically metacresol purple — changes from purple to yellow in the presence of carbon dioxide. It is a disposable qualitative device for confirming tracheal placement when no capnograph is available. It gives no waveform, no number and no trend, and it is unreliable after gastric insufflation or contamination with gastric contents.
  • Mass spectrometry. Sample gas is ionised and the ions are separated by mass-to-charge ratio in a magnetic field. It can identify and quantify every gas in the mixture, is highly accurate, but is expensive and bulky and has largely disappeared from clinical use.
  • Raman spectroscopy. A laser illuminates the sample and a small proportion of scattered light is shifted in wavelength by an amount characteristic of each molecule. It can measure oxygen and nitrogen as well, but the equipment is large and no longer in routine theatre use.
  • Photoacoustic spectroscopy. Pulsed infrared radiation is absorbed by the gas, causing cyclical heating, expansion and a pressure wave that a microphone detects as sound. It is fast, stable and needs no optical detector, and it appears in some multi-gas analysers.
07

The gas infrared cannot see

Oxygen analysis

Oxygen is symmetrical and has no dipole moment, so it does not absorb infrared. Every method of measuring it therefore exploits something else about the molecule: its paramagnetism, or its behaviour at an electrode.
MethodPrincipleStrengthsLimitations
Paramagnetic analysisOxygen has two unpaired electrons in its outer shell and is attracted into a magnetic field. In a differential paramagnetic cell, a rapidly switched magnetic field acting on the sample and a reference gas produces a pressure difference across a transducer, proportional to the difference in oxygen partial pressure.Fast, accurate, no consumable element, continuous breath-by-breath displayWater vapour must be removed; nitric oxide is also paramagnetic and interferes
Fuel cell (galvanic cell)Oxygen diffuses through a membrane to a gold cathode and is reduced; a lead anode is oxidised in potassium hydroxide. The current generated is proportional to the oxygen partial pressure. No external polarising voltage is required.Simple, self-powered, cheap; widely used as a circuit oxygen monitorConsumed in use with a finite life; slow response of roughly 20–30 s; temperature dependent
Polarographic (Clark) electrodeA platinum cathode and a silver/silver chloride anode in potassium chloride, behind an oxygen-permeable membrane, with an external polarising voltage of about 0.6 V. Oxygen is reduced at the cathode (O₂ + 2H₂O + 4e⁻ → 4OH⁻) and the current is proportional to the oxygen partial pressure.Measures PO₂ in blood as well as in gas; the reference method in a blood gas analyserNeeds a polarising voltage and regular calibration; membrane and electrolyte deteriorate; temperature dependent
Pulse oximetryMeasures oxygen saturation of haemoglobin, not partial pressure, from the differential absorption of red and infrared light by oxyhaemoglobin and deoxyhaemoglobin in pulsatile blood.Non-invasive and continuousNot a measure of oxygen tension, content or delivery; unreliable with poor perfusion, dyshaemoglobins and motion
08

Completing the multi-gas analyser

Volatile agent analysis

The same infrared bench that measures carbon dioxide measures the volatile agent, at a different wavelength. Agent identification is a separate problem from agent quantification, and it is solved by measuring at more than one wavelength.
MethodPrincipleCurrent status
Infrared absorptionAll modern volatile agents contain C–H bonds and absorb strongly near 3.3 µm. Measuring at several wavelengths allows the analyser both to quantify the agent and to identify which agent is present.The standard method in current anaesthetic machines
RefractometryThe agent alters the refractive index of the gas mixture, producing an interference pattern.Historically used for calibrating vaporisers rather than for monitoring
Ultraviolet absorptionHalothane absorbs ultraviolet light; other agents do so weakly.Effectively obsolete
Piezoelectric quartz crystalA lipophilic coating on an oscillating quartz crystal absorbs the agent; the added mass lowers the resonant frequency.Cannot distinguish between agents
Mass spectrometry and Raman spectroscopyIonised or scattered light gives a spectrum from which every gas in the mixture can be identified and quantified.Accurate but expensive, bulky and no longer used in routine theatre practice

All the agents in current use are halogenated hydrocarbons or ethers containing C–H bonds, so they all absorb near 3.3 µm. A single-wavelength analyser can therefore tell you how much agent is present but not which agent it is — and since the minimum alveolar concentration of desflurane is roughly fifteen times that of isoflurane, an analyser that misidentifies the agent will report a dangerously misleading depth. Modern analysers sample several wavelengths and match the resulting absorption pattern against known spectra to identify the agent before quantifying it.

09

From measurement to physiology

The physiological effects of hypercarbia

A number on a monitor only matters because of what it does to the patient. Start from the chemistry, then work system by system — and be explicit that the cardiovascular picture is the sum of an indirect sympathetic effect and an opposing direct depressant one.

The chemistry underneath

Carbon dioxide raises hydrogen ion concentration by spontaneous and carbonic anhydrase-catalysed hydration:

Hydration of carbon dioxideCO₂ + H₂O ⇌ H₂CO₃ ⇌ H⁺ + HCO₃⁻

Carbon dioxide also reacts with free amine groups in proteins to form carbamate compounds. At cellular level it diffuses readily across cell membranes and generates carbonic acid and hydrogen ions inside the cell, but intracellular buffering is rapid, reaching about 90% completion within three hours of the onset of hypercarbia. The resulting pH follows the Henderson–Hasselbalch relationship:

Henderson–HasselbalchpH = 6.1 + log₁₀ ( [HCO₃⁻] ÷ (0.03 × PaCO₂ in mmHg) )

As rules of thumb for the exam: in acute respiratory acidosis pH falls by approximately 0.08 for every 10 mmHg (1.33 kPa) rise in PaCO₂, and bicarbonate rises by about 1 mmol/L. With chronic hypercarbia, renal compensation raises bicarbonate by about 4 mmol/L for each 10 mmHg, over two to three days.

System by system

SystemMechanismEffectQualifier worth stating
CardiovascularIncreased sympathetic discharge with raised plasma adrenaline and noradrenalineIncreased contractility, cardiac output, heart rate and blood pressure; increased risk of arrhythmias; systemic vasodilatation with warm flushed peripheries and a bounding pulseDirect effects of carbon dioxide and acidaemia on the myocardium and vascular smooth muscle are depressant and vasodilatory. At severe levels the direct depression dominates and cardiac output falls.
Central nervous systemCerebral vasodilatation; direct narcotic effect at high tensionsIncreased cerebral blood flow and intracranial pressure, lowered seizure threshold, headache, confusion, drowsiness and eventually carbon dioxide narcosisCarbon dioxide crosses the blood–brain barrier freely whereas bicarbonate does not, so acute changes act rapidly on the central chemoreceptors.
RespiratoryStimulation of central and peripheral chemoreceptorsIncreased minute ventilation, mainly through tidal volume; bronchodilatation; enhanced hypoxic pulmonary vasoconstriction with improved ventilation–perfusion matching; increased pulmonary vascular resistanceVolatile agents flatten the ventilatory response to carbon dioxide and shift the apnoeic threshold, which is why hypercarbia is common under anaesthesia.
Oxygen transportBohr effectRightward shift of the oxyhaemoglobin dissociation curve, reducing haemoglobin affinity and favouring tissue unloading; the Haldane effect assists carbon dioxide offloading in the lungIf the inspired oxygen fraction is fixed, a rising PaCO₂ lowers alveolar PO₂ through the alveolar gas equation.
Metabolic and renalH⁺/K⁺ exchange; renal compensation over daysHyperkalaemia; increased ionised calcium; increased renal bicarbonate reabsorption and hydrogen ion excretion over two to three daysIntracellular buffering is rapid, reaching about 90% completion within three hours of the onset of hypercarbia.
OtherImmunomodulatory effects of hypercapnic acidosisDampening of neutrophil and cytokine-mediated lung injury; sweating, agitation and muscle twitching; increased uterine blood flowThese protective effects are part of the rationale for permissive hypercapnia in acute respiratory distress syndrome.

And the mirror image: hypocarbia

  • Cerebral vasoconstriction with reduced cerebral blood flow — useful briefly to reduce intracranial pressure, harmful if sustained because it risks cerebral ischaemia. This is why routine prolonged hyperventilation is no longer recommended after head injury.
  • Leftward shift of the oxyhaemoglobin dissociation curve, increasing haemoglobin affinity and impairing tissue unloading.
  • Hypokalaemia and reduced ionised calcium, with paraesthesiae, tetany and a positive Chvostek sign in the awake patient.
  • Reduced cardiac output at extremes, coronary vasoconstriction and arrhythmias.
  • Apnoea in the anaesthetised patient once PaCO₂ falls below the apnoeic threshold.
  • Inhibition of hypoxic pulmonary vasoconstriction, worsening ventilation–perfusion matching.
10

Reading the trace

Drawing a capnogram, and what has to be on it

A capnogram carries almost nothing without its labelling: the same four phases describe a normal trace, an obstructed one and a disconnection, and it is the axes, the units and the angles that separate them.

The drawing checklist

The diagram itself
  • y-axis: PCO₂ in mmHg or kPa, with the unit written
  • x-axis: time in seconds
  • Baseline sitting on zero
  • All four phases numbered
  • End-tidal point marked at the end of phase III
  • Expiration and inspiration indicated
  • α and β angles labelled with their normal values
The prose beside it
  • What gas each phase represents
  • Why phase III slopes slightly upward
  • Normal ETCO₂ in both units
  • The arterial to end-tidal gradient, and what widens it
  • One sentence on what each angle indicates

The three past questions this topic has been set from, with their answers worked in full, are in the question bank: defining and measuring end-tidal carbon dioxide, how a capnometer measures it and the types of capnography and their role in practice.

Connecting…
Account progress

Connecting your study progress…

Account & profile